ALCON RESEARCH, LTD. - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: fourteen (14) unopened pouches of lot number 32560021 were returned for evaluation (transported in an opened carton lot number: 32552376.Samples were evaluated.There were no damaged or broken cartridges observed.A photo of the side of the monarch carton was provided.Product history records were reviewed and the documentation indicated the product met release criteria.The customer indicated the use of a non-qualified lens model and non-qualified viscoelastic.The lens model indicated is only approved for use in the company b and c cartridges with only approved viscoelastic.The iol product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.Root cause: the reported damage could not be verified because the used company (d) cartridge was not returned.The root cause for the reported damage is most likely related a failure to follow the dfu.The account used a lens/monarch combination which is not qualified for use.The use of non-qualified combinations may result in delivery issues and/or damage.In addition, the viscoelastic indicated is not qualified for the associated lens model with qualified company combinations.Due to differing material properties, the use of a non-qualified viscoelastic may result in delivery issues and/or damage.A photo is attached of the side of the company d carton.Information in the file attachment indicated what appears to be a customer misunderstanding involving an assumption that there is no company a lens model, and that the company b lens model listed as an approved lens model must refer to the company c (which was used in the cartridge); however, the company b lens model is a valid lens model with different dimensions than the company c lens model.The company b has been qualified for use in the company c cartridge.The lens model indicated as used by the customer is an company b lens model and is not qualified for use in the company d cartridge.The company b lens model is only approved for use in the company b and c cartridges with only approved viscoelastic.The manufacturer internal reference number is: (b)(4).
 
Event Description
A doctor of ophthalmology reported that a cartridge broke during the intraocular lens implanting process.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7465678
MDR Text Key106847917
Report Number1119421-2018-00474
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Catalogue Number8065977763
Device Lot Number32560021
Other Device ID Number380659777639
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2018
Date Device Manufactured08/16/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-