Model Number 383400
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 03/20/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that an issue occurred with an enteral feeding pump.No patient was involved.Upon follow up with the customer on (b)(6) 2018 the customer stated that the screen lit up but no information was displayed.This was not due to physical damage.
 
Manufacturer Narrative
The unit was triaged and the customer¿s reported condition was confirmed.Liquid ingress caused the pcba to corrode, and is the result of customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7466721
MDR Text Key106711796
Report Number3006451981-2018-00312
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date01/25/2015
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2018
Initial Date Manufacturer Received 04/25/2018
Initial Date FDA Received04/26/2018
Supplement Dates Manufacturer Received04/25/2018
Supplement Dates FDA Received09/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-