Model Number N/A
Device Problems Bent (1059); Device Damaged by Another Device (2915)
Patient Problem No Information (3190)
Event Date 03/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant products: ep-115393, e1 44-36 std hmrl brng, 963040; ep-115393, e1 44-36 std hmrl brng, 654320.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2018 - 02892, 0001825034 - 2018 - 02893.
 
Event Description
It was reported that the first humeral bearing was damaged during a back table assembly into the humeral tray.A new ring was put into the humeral tray.The second humeral bearing was also damaged during back table assembly.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the ring loc shows that it is bent and damaged.Dimensions taken on ring loc are conforming to specifications.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RINGLOC+ REPLACEMENT RING SZ21
Type of Device
PROTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7467734
MDR Text Key106736406
Report Number0001825034-2018-02891
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK070369
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number106021
Device Lot Number039870
Other Device ID Number(01) 00880304355177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/04/2018
Initial Date FDA Received04/27/2018
Supplement Dates Manufacturer Received07/02/2018
Supplement Dates FDA Received07/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-