Catalog Number 500101
Device Problems Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Bacterial Infection (1735); Death (1802); Hemorrhage/Bleeding (1888); Neurological Deficit/Dysfunction (1982)
Event Type  Death  
Manufacturer Narrative
Investigation summary: with a review of the available information, there is no evidence of a device malfunction or performance issues that would impact the reported events.Possible clinical factors that may have contributed to this event include the patients pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulants, antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.This device is used for treatment, not diagnosis.The syncardia 70cc temporary total artificial heart (tah-t) is an implantable pulsatile biventricular replacement device that replaces a patient's native ventricles and valves and pumps blood to both the pulmonary and systemic circulation systems.The syncardia 70cc tah-t is indicated for use as a bridge to transplantation in cardiac transplant-eligible candidates at risk of imminent death from biventricular failure.The syncardia tah-t system is intended for use inside and outside the hospital.This event was assessed and is being reported as part of a retrospective review of events in response to an update to the mdr complaint sources.If new or additional information is received in the future, syncardia will file a follow-up mdr.(b)(4).
 
Event Description
A retrospective review of intermacs de-identified data collected between (b)(6) 2015 and (b)(6) 2017 revealed that a patient expired with a cause of death listed as device malfunction.The intermacs data listed that the tah-t did not function normally.The intermacs data listed that it was unknown if the tah-t was explanted.
 
Manufacturer Narrative
(b)(4).Follow-up report 1.
 
Event Description
While supported by the tah-t for 504 days, the patient experienced the following adverse events as defined by intermacs: 29 days post implant - infection/location: positive blood cultures/type: bacterial; 32 days post implant - infection/location: other/type: bacterial; 35 days post implant - neurological dysfunction - cva/location: left hemisphere: occipital/severity: other; 120 days post implant - bleeding; 132 days post implant - bleeding; 189 days post implant - device malfunction and/or pump thrombosis; 205 days post implant - infection/location: positive blood cultures/type: bacterial; 205 days post implant - other sae; 218 days post implant - psychiatric episode; 312 days post implant - device malfunction and/or pump thrombosis; 504 days post implant - device malfunction and/or pump thrombosis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TAH-T
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key7468858
MDR Text Key106757009
Report Number3003761017-2018-00193
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Catalogue Number500101
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/28/2017
Initial Date FDA Received04/27/2018
Supplement Dates Manufacturer Received11/28/2017
Supplement Dates FDA Received07/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Weight70
-
-