Model Number DL-344-10-K
Device Problems Signal Artifact/Noise (1036); Break (1069); High impedance (1291)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/15/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Review of the patient's impedance and ecog data was performed, the data is consistent with an intermittent lead break.The explanted lead was not returned to neuropace for analysis, however review of the lead impedance and ecog data was performed and the data is consistent with an lead break.Unable to determine root cause.
 
Event Description
The neurologist observed a signal artifact on the ecog signal from the patient's lead connected to port 1 of the neurostimulator, indicating a potential lead break.The artifact was observed/recreated while running a real-time ecog and palpating around the region of the neurostimulator and where the lead was secured via the dog bone plate.The physician then programmed the neurostimulator so the affected lead was no longer part of the therapy pathway.On (b)(6) 2018, a lead revision was performed.Detection and therapy has been restored to both leads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key7470071
MDR Text Key106794583
Report Number3004426659-2018-00013
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005014
UDI-Public010085554700501417190720
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberDL-344-10-K
Device Catalogue Number1007607
Device Lot Number22420-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age53 YR
-
-