(b)(4).Review of the patient's impedance and ecog data was performed, the data is consistent with an intermittent lead break.The explanted lead was not returned to neuropace for analysis, however review of the lead impedance and ecog data was performed and the data is consistent with an lead break.Unable to determine root cause.
|
The neurologist observed a signal artifact on the ecog signal from the patient's lead connected to port 1 of the neurostimulator, indicating a potential lead break.The artifact was observed/recreated while running a real-time ecog and palpating around the region of the neurostimulator and where the lead was secured via the dog bone plate.The physician then programmed the neurostimulator so the affected lead was no longer part of the therapy pathway.On (b)(6) 2018, a lead revision was performed.Detection and therapy has been restored to both leads.
|