MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/02/2018
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by manufacturer: on (b)(4) 2018, a field service engineer (fse) followed-up over-the-phone and instructed the customer to clean the sampling needle guide rod and run patient samples.The customer had also replaced the sampling needle assembly on (b)(6) 2018, the customer reported that after cleaning the sampling needle guide rod quality controls and patient samples were run without any issues.The customer reported that it was likely that the issue occurred due the sampling needle assembly being bent in addition to the dirty guide rod.The customer reported that no further issues were being experienced with the g8 instrument.No further action was required by the fse.A (b)(6) complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There were no other similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: the 6.3 error messages when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.The 710 z1-axis error an abnormality occurred in the up and down movement of the sampling needle.If this occurs during a stat assay, check that the container setting (cup or tube) is correctly set.The error also occurs when the sample vial was not recognized as a primary tube, due to the disoriented sample sensor.The most probable cause of the reported event was due to a bent sampling needle assembly and dirty guide rod.(b)(4).
 
Event Description
On (b)(6) 2018, a customer reported getting 710 z-axis error messages on the g8 analyzer.The technical support specialist (tss) instructed to remove the brown cover from the analyzer and clean and lubricate the sampling needle assembly guide rods.The customer then proceeded to run five (5) patient samples without any errors.On (b)(6) 2018, the customer reported that the 710 z-axis error message persisted on the g8 analyzer.The tss sent a sampling needle assembly to the customer for replacement.On (b)(6) 2018, the customer reported that after replacing the sampling needle assembly the 710 z-axis error message persisted.The a field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 105-8 623
JA  105-8623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7471056
MDR Text Key107211180
Report Number8031673-2018-00365
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/27/2018
Distributor Facility Aware Date04/02/2018
Device Age7 YR
Event Location Hospital
Date Report to Manufacturer04/27/2018
Date Manufacturer Received04/02/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-