DEPUY SYNTHES PRODUCTS LLC K-WIRE ATTACHMENT FOR PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.037
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that the k-wire attachment device did not function and parts of the device were coming off.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the bearing of the attachment device seized, jammed, and was moving heavily.It was noted that the attachment was rough running.It was further determined that the device failed pretest for function test, check of free moving, check clamping rang on tool coupling, and check sleeve for spring tick ¿pass¿, if hexagon end.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
K-WIRE ATTACHMENT FOR PEN DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7473532
MDR Text Key106978408
Report Number8030965-2018-53449
Device Sequence Number1
Product Code HWE
UDI-Device Identifier7611819167872
UDI-Public(01)7611819167872
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.037
Device Lot Number7697649
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-