Catalog Number ASKU
Device Problems Hole In Material (1293); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/01/2018
Event Type  malfunction  
Manufacturer Narrative
An alarm indicative of a potential malfunction of the disposable cassette was identified.The device was not returned and the lot number was unknown; therefore, a device analysis could not be completed.However, a hole in the tubing of a homechoice cassette was reported and is known to cause this alarm.The cause of the hole could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a system error 2240 alarm (air in set/line) occurred on a homechoice device during peritoneal dialysis therapy.This event occurred during drain three of five.The patient was connected at the time of the alarm.The patient stated that there was hole in the patient line tubing.The patient also stated that there were no pets but saw a rat a few days ago.The renal therapy service agent (rtsa) had the patient cycle power on the homechoice.The rtsa assisted the patient in proper disconnect procedures.The rtsa reviewed proper procedures with the customer.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7473835
MDR Text Key106928811
Report Number1416980-2018-02485
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received04/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
HOMECHOICE
Patient Age34 YR
-
-