Boston scientific received information that the cardiac resynchronization therapy defibrillator (crt-d) and another manufacturer¿s high voltage right ventricular (rv) lead exhibited high out of range shock impedance measurements.During the in-office interrogation the high shock impedance was unable to be reproduced.Further evaluation of the crt-d system found that another manufacturer¿s right atrial (ra) lead impedance measurements were variable with evidence of oversensing noise from the minute ventilation signal.Subsequently the rate response trend was programmed off.Boston scientific technical services (ts) was contacted and reviewed the data stored in the remote home monitoring system.It was found that the shock impedance increased from 40 ohms to 60 ohms and then to greater than 200 ohms.No sources of electromagnetic interference (emi) could be identified.The patient was seen again three weeks later where the increase in ra lead impedance measurements were able to be recreated with pocket manipulation.It was noted when reprogrammed from triad to rv coil to can there were no high shock impedance measurements.It was further noted that another manufacturer¿s left ventricular (lv) lead exhibited spikes in impedance measurements that were unable to be reproduced.The clinic noted that this had been occurring for some time and had been programmed rv to lv tip, a possible connection issue between the crt-d and lv lead spring contact was considered as a potential cause.This connection issue was also thought to be a potential cause for the ra lead noise and impedance changes.The high shock impedance measurements were thought to be caused by possible growth on the proximal coil or a lead fracture.The clinic note that the patient has not had any tachycardia therapy since the crt-d was implanted, thus they are considering downgrading the patient to a cardiac resynchronization therapy pacemaker (crt-p) device.At this time the clinic has opted to continue to monitor.
|