(b)(4).Device evaluation by manufacturer: a (b)(4) complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There were no other similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 1, introduction and applications, section 1.6 - procedures provides detailed information on column preparation and installation.In chapter 5: maintenance procedures, section 5.6 - column replacement provides detailed information on how to replace the column.The most probable cause of the reported event was due to a faulty column.
|
On (b)(6) 2018, a customer reported getting calibration errors on the g8 instrument after replacing the column on (b)(6) 2018.Customer stated she primed and ran calibrators and the calibration failed.The customer made a second attempt and the calibration failed again.Customer tried to use their second instrument on site, switched the column, attempted calibrating and calibration failed.Customer had another new column, replaced column a second time and was able to calibrate and run controls successfully, all results were in acceptable range.This caused a delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.Technical support specialist sent a replacement column to customer.The customer confirmed receipt of the new column on (b)(6) 2018.No further issues were reported.The g8 instrument is functioning as expected.
|