• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1054096
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
Device has been repaired but not returned to the customer, pending customer approval of estimate.
 
Event Description
The manufacturer received information alleging a ventilator failed a test step during testing.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The device was recalibrated to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin dr.
new kensington, PA 15068
7243349303
MDR Report Key7479207
MDR Text Key107145923
Report Number2518422-2018-00991
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959022744
UDI-Public00606959022744
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1054096
Device Catalogue Number1054096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/03/2018
Initial Date FDA Received05/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-