• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ORANGE ENTERAL INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY ORANGE ENTERAL INFUSION PUMP Back to Search Results
Model Number INFORNG
Device Problem Failure to Auto Stop (2938)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/06/2018
Event Type  malfunction  
Manufacturer Narrative
This device was returned to mmdg for evaluation.The pump operated within product specifications.The pump history was also reviewed where the pump shows that it had alarmed no food multiple times before it was turned off.Mmdg could not confirm this complaint.There is no indication that the complaint occurred as reported.This report is being filed because the event occurred while the device was in use by a pediatric patient.Per our procedures all reports of this type of event that involve a pediatric patient are considered reportable.
 
Event Description
The initial reporter stated that the pump continues to pump air after the bag is empty.The initial reporter also stated that they had seen air in the tubing between the pump and the patient and stopped the pump before air reached the patient.They stated that they also tested the pump and that the pump pumped air at that time as well.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY ORANGE ENTERAL INFUSION PUMP
Type of Device
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city UT 84123
Manufacturer Contact
kristin hardesty
4314 zevex park lane
salt lake city, UT 84123
8012641001
MDR Report Key7480348
MDR Text Key107196953
Report Number1722139-2018-00107
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberINFORNG
Device Catalogue NumberINFORNG
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/06/2018
Initial Date FDA Received05/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
-
-