MAKO SURGICAL CORP. TIBIAL INLAY IMPACTOR HEAD; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLY
|
Back to Search Results |
|
Catalog Number 160177 |
Device Problems
Break (1069); Fracture (1260)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
|
Event Description
|
Impactor broke during impaction of tibia pka implantation.
|
|
Manufacturer Narrative
|
An event regarding crack/fracture involving a mako impactor was reported.The event was not confirmed.Method & results: product evaluation and results: visual, dimensional, functional inspection, and material analysis were not performed as the item was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: a review of the device history records could not be performed as no lot information was provided.Complaint history review: a complaint history review could not be performed as no lot information was provided.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient medical information was provided.Further information such as returned devices are needed.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
|
|
Event Description
|
Impactor broke during impaction of tibia pka implantation.Update: the blue impactor tip broke during impaction.
|
|