(b)(4).Device evaluation by manufacturer: a field service engineer (fse) was dispatched to the customer's facility to address the reported event.Fse confirmed the reported issue with the error printout from the g8 instrument.Fse found the large syringe guide rods dry and binding.Fse lubricated the guide rods, lead screw, small syringe, and the sampling unit.Fse then ran quality controls (qc) samples; all results were with acceptable range.The error cleared and no grinding noise is heard.The g8 instrument is performing as intended.No further action was required from field service.A 13-month complaint history review and service history review for serial (b)(4) from 09-mar-2017 through the awareness date (b)(6) 2018 for similar complaints was performed.There were two (2x) other similar complaints identified during the searched period.The g8 operator's manual under chapter 6 - troubleshooting, states message 706 is syringe-l error and is generated due to operation error in syringe-l.The countermeasure is for the user to inspect the syringe-l and execute smp.Reset.Also under chapter 5: maintenance procedures outlines and provides guidance for instrument care and maintenance.The customer is to call tosoh support should questions arise.The most probable cause of the reported event was due to the large syringe needed to be lubricated.
|