ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-04
Device Problem Extrusion (2934)
Patient Problem No Code Available (3191)
Event Date 04/11/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
On (b)(6) 2017, the recipient reportedly experienced device extrusion.The recipient underwent skin flap surgery followed by a round of antibiotics.On (b)(6) 2017, the recipient's device extruded once again.On (b)(6) 2017, the recipient's device was explanted.
 
Manufacturer Narrative
The recipient will reportedly not be reimplanted.The device passed the external visual and photographic imaging inspections.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
delilah garcia
28515 westinghouse place
valencia, CA 91355
6613621741
MDR Report Key7485309
MDR Text Key107313606
Report Number3006556115-2018-00202
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Audiologist
Type of Report Initial,Followup,Followup
Report Date 04/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date07/31/2018
Device Model NumberCI-1500-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/03/2018
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received07/05/2018
08/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
-
-