• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TULSA DENTAL PRODUCTS LLC PROTAPER GOLD ROTARY F1 21MM; DRILL, DENTAL, INTRAORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TULSA DENTAL PRODUCTS LLC PROTAPER GOLD ROTARY F1 21MM; DRILL, DENTAL, INTRAORAL Back to Search Results
Catalog Number PTGRF121
Device Problem Material Too Rigid or Stiff (1544)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/07/2018
Event Type  malfunction  
Manufacturer Narrative
As a result of this error, the files lack flexibility, which can lead to file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.
 
Event Description
It was reported that a protaper gold file was not bending like usual.The doctor stated the files felt more like protaper universal files.However, upon investigation on april 6, 2018, it was found that the files were in fact protaper gold files, but they were not heat treated to specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTAPER GOLD ROTARY F1 21MM
Type of Device
DRILL, DENTAL, INTRAORAL
Manufacturer (Section D)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7486766
MDR Text Key107344444
Report Number2320721-2018-00035
Device Sequence Number1
Product Code DZA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Remedial Action Recall
Type of Report Initial
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPTGRF121
Device Lot Number00173645
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/06/2018
Initial Date FDA Received05/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2018
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-