TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 816300 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/17/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) no consequences or impact to patient.(b)(4) device operates differently than expected.(b)(4) monitor.Evaluation is in progress, but not yet concluded.Per the user facility's biomedical technician, the cursor would move up and down but not left and right.He checked and reset the cable and that temporarily fixed the cursor problem.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) cursor would not respond to touch.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the cursor to not respond to touch and would not move from right edge of screen.The central control monitor (ccm) was calibrated using a universal serial bus (usb) mouse and key board and the cursor would respond to touch.However, it could not be calibrated correctly with the buttons.It was determined that the touch screen was the cause of the problem.The service repair technician (srt) replaced the touch screen.The unit operated to the manufacturer's specifications.This complaint is related to (b)(4) / medwatch #1828100-2018-00246.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|