• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 816300
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4) no consequences or impact to patient.(b)(4) device operates differently than expected.(b)(4) monitor.Evaluation is in progress, but not yet concluded.Per the user facility's biomedical technician, the cursor would move up and down but not left and right.He checked and reset the cable and that temporarily fixed the cursor problem.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the central control monitor (ccm) cursor would not respond to touch.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the cursor to not respond to touch and would not move from right edge of screen.The central control monitor (ccm) was calibrated using a universal serial bus (usb) mouse and key board and the cursor would respond to touch.However, it could not be calibrated correctly with the buttons.It was determined that the touch screen was the cause of the problem.The service repair technician (srt) replaced the touch screen.The unit operated to the manufacturer's specifications.This complaint is related to (b)(4) / medwatch #1828100-2018-00246.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7488305
MDR Text Key107469290
Report Number1828100-2018-00244
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K163531
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number816300
Device Catalogue Number816300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/17/2018
Initial Date FDA Received05/04/2018
Supplement Dates Manufacturer Received06/21/2018
Supplement Dates FDA Received06/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number1828100-10/14/2011-020-C
Patient Sequence Number1
-
-