• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Patient Involvement (2645)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
The customer reported complaint for the platform displayed an error message user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) was confirmed during archive data review and functional testing.Load cells were found defective and a replacement was required.The report of a user advisory (ua) 12 (lifeband not present) was not confirmed during functional testing.Platform displayed the (ua) 12 error message upon powering on the platform without the lifeband being inserted, which the platform is functioning as intended.Visual inspection was performed and found cracked front enclosure and a bent battery lock, unrelated to the reported complaint.Front enclosure and battery lock were replaced to address the issue.The autopulse platform is a reusable device as well as a serviceable device and was manufactured in 2008.Therefore, this type of physical damage found during visual inspection is characteristic of normal wear and tear for the life of the device.The platform has also exceeded its expected service life of 5 years.Following the repair, the platform was further tested for 15 minutes with large resuscitation testing fixture, (lrtf) equivalent to 250 pound patient and passed all functional testing criteria and met all required specifications.During functional testing, (ua) 07 error message displayed upon powered on the device, thus confirming the reported complaint.Review of the archive data indicated error messages (ua) 07 on the customer reported event date.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse with serial number (b)(4).
 
Event Description
During shift check, the autopulse platform ((b)(4)) was displaying error messages user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) and (ua) 12 (lifeband not present).Customer was unable to clear the errors.No patient involvement was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
noemi schambach
2000 ringwood ave,
san jose, CA 95131
4084192955
MDR Report Key7488993
MDR Text Key107606077
Report Number3010617000-2018-00515
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000277
UDI-Public00849111000277
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-