ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problem Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/19/2018
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported during an intraocular lens (iol) implant procedure using a preloaded device, one of the haptics was torn off.A backup lens was used to complete the procedure.
 
Manufacturer Narrative
The product was not returned.Photos were available which show the back of a carton visible with the device.The device is on top of what appears to be a envelope.The device in the photo is shown on its side.The mylar is visible with the diopter shown (23.5).The lens stop and plunger lock have been removed.The lens door is closed.It cannot be determined from the photo if viscoelastic is present in the device.The plunger appears to have been retracted into the loading area.There appears to be a lens or lens material present in or on the nozzle.The photo was viewed at varying magnifications.Due to the blurriness of the photo, the reported lens damage cannot be verified.Product history records were reviewed and the documentation indicated the product met release criteria.Viscoelastic was not provided.It is unknown if a qualified product was used.Based on our observation of the attached photos, the reported haptic damage cannot be confirmed.There appears to be a lens advanced into the nozzle and possibly lens material on the nozzle of the device.The photo is too blurry upon higher magnification to confirm.It is difficult to make a determination of handling / damage without evaluation of the physical sample.A final root cause cannot be determined based on available information.It is unknown if a qualified viscoelastic was used.Material properties of non-qualified viscoelastics may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7491096
MDR Text Key107600324
Report Number1119421-2018-00538
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeBO
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Model NumberAU00T0
Device Lot Number12575597
Other Device ID Number00380652358347
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-