SMITHS MEDICAL, ASD, INC. JELCO® VIAVALVE® SAFETY IV CATHETER; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
Catalog Number 326310 |
Device Problem
Fail-Safe Design Failure (1222)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
It was reported that when the needle of a jelco® viavalve® safety iv catheter was inserted and hub removed, blood started.No report of patient injury.
|
|
Manufacturer Narrative
|
One used peripheral intravenous catheter was returned for analysis with a locked out needle guard sub-assembly.The catheter sub-assembly was not included so the valve could not be evaluated.The sample had blood residue in the flash chamber, nose and guard.Blood present in the guard is indicative that the deployment chassis was left attached to the catheter hub for a longer than necessary time.The beveled acam which houses the seal was not returned for evaluation.T would appear that the excessive presence of blood may be a deployment training issue as the deployment chassis would have to be left attached to the catheter hub for a long period of time for the amount of blood to leak.Based on these observations and the lack of a beveled catheter assembly to examine, the complaint cannot be confirmed.
|
|