• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number L5C4531
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/12/2018
Event Type  malfunction  
Manufacturer Narrative
The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that air was noticed in the tubing of a homechoice cassette without an alarm during peritoneal dialysis therapy.This event occurred during dwell one of three while the patient was connected.The renal therapy service agent (rtsa) had the patient close transfer set.The patient stated that there was no leak.The rtsa had the patient check the connections to the two bags.The patient stated that the connections were tight and the blue clamp was closed.The rtsa assisted patient to end therapy with the removal of the cassette.The patient stated that they can see bubbles in the cassette but nothing is leaking.There was nothing found during trouble shooting that could have caused or contributed to the reported event.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7497066
MDR Text Key107735579
Report Number1416980-2018-02712
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00085412090078
UDI-Public(01)00085412090078
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/17/2023
Device Catalogue NumberL5C4531
Device Lot NumberH18A17044
Was Device Available for Evaluation? No
Date Manufacturer Received04/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age69 YR
-
-