• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Patient Involvement (2645)
Event Date 04/10/2018
Event Type  malfunction  
Manufacturer Narrative
The customer reported complaint for the platform displayed an error message user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) was confirmed during archive data review and functional testing.Both load cells were found defective and a replacement was required.Visual inspection was performed and found front and bottom enclosures, bent battery springs and a top cover with signs of corrosion, unrelated to the reported complaint.Front and bottom enclosures, bent battery springs and a top cover were replaced to address the issue.Following the repair, the platform passed all functional testing criteria and met all required specifications.Review of the archive data indicated multiple error messages (ua) 07 on the customer reported event date.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse with serial number (b)(4).
 
Event Description
During a shift check, the autopulse platform ((b)(4)) displayed a user advisory 7 (discrepancy between load 1 and load 2 too large) message.The user was unable to clear the error message.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
noemi schambach
2000 ringwood ave,
san jose, CA 95131
4084192955
MDR Report Key7499515
MDR Text Key107877718
Report Number3010617000-2018-00537
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/10/2018
Initial Date FDA Received05/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-