• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TULSA DENTAL PRODUCTS LLC X-SMART IQ MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TULSA DENTAL PRODUCTS LLC X-SMART IQ MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number XSIQEM1HP
Device Problems Retraction Problem (1536); Device Operates Differently Than Expected (2913)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
It was reported that an x-smart iq motor was auto-reversing improperly.The event outcome is unknown as of this mdr evaluation.
 
Manufacturer Narrative
Multiple unsuccessful attempts were made to obtain the device for evaluation.Multiple unsuccessful attempts were made to obtain the patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART IQ MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
MDR Report Key7502748
MDR Text Key107985584
Report Number2320721-2018-00037
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
PMA/PMN Number
K161213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberXSIQEM1HP
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-