• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383400
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An evaluation of the kangaroo pump was performed and the customer states, ¿pump shuts off when pushing fourth button".The unit was triaged and the reported issue could not be confirmed at this time; turned on unit and pressed all the buttons to try and duplicate the complaint.Set the settings to 400 ml/hr and let the unit run for about thirty minutes while pressing the buttons occasionally to try and duplicate the customer complaint.Could not duplicate the complaint.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer returned the enteral feeding pump without prior notification.During triage of the pump on (b)(6) 2018, the service tech found a note from the customer attached to the unit that stated the pump shuts off when pushing fourth button.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire st
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire st
mansfield MA 02048
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7503501
MDR Text Key107961265
Report Number1282497-2018-00351
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006645
UDI-Public10884521006645
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2018
Initial Date Manufacturer Received 04/13/2018
Initial Date FDA Received05/10/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-