Peritoneal dialysis patient report provided treatment to technical support during troubleshooting of the liberty select.The treatment data was assessed for 150% and 180% iipv and did reveal an episode of increased intraperitoneal volume (iipv).The largest drain volume from this treatment occurred during drain 3 where 3284ml drained.This drain is 187% of the prescribed fill volume.In order to be a reportable malfunction, at least 180% of the prescribed fill volume would have to have been drained.180% of the prescribed fill volume of 1753ml is 3155ml.Therefore, a reportable malfunction has occurred.The patient was advised to discontinue use of the liberty select and to contact the clinic nurse.The patient reported that since there is no continuous ambulatory peritoneal dialysis supplies, that the patient would continue use of the cycler until a replacement was delivered.Additional information was solicited.The liberty select was replaced.
|