CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Model Number LIBERTY SELECT
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2018
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the plant¿s investigation.
 
Event Description
Peritoneal dialysis patient report provided treatment to technical support during troubleshooting of the liberty select.The treatment data was assessed for 150% and 180% iipv and did reveal an episode of increased intraperitoneal volume (iipv).The largest drain volume from this treatment occurred during drain 3 where 3284ml drained.This drain is 187% of the prescribed fill volume.In order to be a reportable malfunction, at least 180% of the prescribed fill volume would have to have been drained.180% of the prescribed fill volume of 1753ml is 3155ml.Therefore, a reportable malfunction has occurred.The patient was advised to discontinue use of the liberty select and to contact the clinic nurse.The patient reported that since there is no continuous ambulatory peritoneal dialysis supplies, that the patient would continue use of the cycler until a replacement was delivered.Additional information was solicited.The liberty select was replaced.
 
Event Description
".".
 
Manufacturer Narrative
The reported complaint symptom (dc drain complication encountered) was not confirmed.The reported complaint symptom (iipv (dv > 180% to 200% of fv) was not confirmed.The reported complaint symptom (scale reading error warning) was not confirmed.A visual inspection of the returned cycler exterior showed no signs of physical damage.An (as-received) simulated treatment was performed and completed without failures.The cycler weighed fill volume values were within tolerance for a liberty cycler.The system air leak test passed.The valve actuation test passed.The load cell verification was within tolerance.There were no discrepancies encountered in the internal inspection of the cycler.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key7504047
MDR Text Key108146660
Report Number2937457-2018-01306
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
PMA/PMN Number
K171652
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLIBERTY SELECT
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2018
Was the Report Sent to FDA? No
Device AgeMO
Date Manufacturer Received06/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
LIBERTY CYCLER SET; PD SOLUTION
-
-