• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR LTD NUCLEUS CI24RE (CA) COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR LTD NUCLEUS CI24RE (CA) COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE; NUCLEUS 24 COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number CI24RE(CA)
Device Problem Extrusion (2934)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
This report is submitted on may 11, 2018.
 
Event Description
It was reported that the patient experienced extrusion of the electrode resulting in the decision to explant the device on (b)(6) 2018.The patient was reimplanted with another device during the same surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI24RE (CA) COCHLEAR IMPLANT WITH CONTOUR ADVANCE ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS  2109
Manufacturer Contact
sujeewa wijesinghe
1 university avenue
macquarie university, nsw 2109
AS   2109
MDR Report Key7505678
MDR Text Key107957629
Report Number6000034-2018-00978
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502007320
UDI-Public(01)09321502007320(11)120313(17)140312
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P970051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Audiologist
Type of Report Initial
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/12/2014
Device Model NumberCI24RE(CA)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/20/2018
Initial Date FDA Received05/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-