• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA UNI MED TIB; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA UNI MED TIB; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problem Bone Fracture(s) (1870)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Report source: japan.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Product remains implanted.
 
Event Description
It was reported that the patient suffered non-displaced tibial plateau fracture 2 weeks post op from vertical cut which was treated with plate and screws.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
(b)(4).Reported event was confirmed by review of x-rays provided.Early postop x-ray shows a fracture at the medial cortex of the medial tibial plateau in vicinity of a previously visualized osteophyte.X-rays obtained 2 weeks postoperatively show a vertical component of fracture of the medial tibial plateau, at the medial margin of the tibial component.Device history record was reviewed and no discrepancies were found.No compatibility issues were noted.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA UNI MED TIB
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7512607
MDR Text Key108225792
Report Number0001825034-2018-03308
Device Sequence Number1
Product Code HSX
Combination Product (y/n)N
PMA/PMN Number
K161592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 01/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number42538000401
Device Lot Number63687978
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-