MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; IRRIGATION SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SYRINGE IRRIGATION BULK / CLEAN, 50ML; IRRIGATION SYRINGE Back to Search Results
Catalog Number 0935280
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that foreign material was found inside the shaft of the syringe.
 
Manufacturer Narrative
Received 5 unused syringes for evaluation.The reported event was confirmed as a supplier related issue.During the visual evaluation, it was noticed that all syringes had foreign material in them.For sample 1, 2, 3 and 4, the foreign material was embedded and exceeded its specification.The sample 5 had a hair attached with tape.The samples were received out of their packages.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "irrigation syringe bulb type non sterile.".
 
Event Description
It was reported that foreign material was found inside the shaft of the syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE IRRIGATION BULK / CLEAN, 50ML
Type of Device
IRRIGATION SYRINGE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7514500
MDR Text Key108411741
Report Number1018233-2018-01735
Device Sequence Number1
Product Code KYZ
UDI-Device Identifier10801741051378
UDI-Public(01)10801741051378
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0935280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Initial Date Manufacturer Received 04/23/2018
Initial Date FDA Received05/15/2018
Supplement Dates Manufacturer Received06/25/2018
Supplement Dates FDA Received07/02/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-