Model Number 2110-0500 |
Device Problems
Hole In Material (1293); Leak/Splash (1354)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
|
Event Description
|
The customer reported that after 1 min of infusing solumedrol, the user observed drops dripping from the pump module.The user opened the door and noted fluid squirting out of tubing.No patient harm was reported.
|
|
Manufacturer Narrative
|
The customer¿s report of leaking at the top of the silicone segment was confirmed.Visual inspection of the set showed that the silicone segment had a tear near the upper fitment measuring 0.0368 inches long.Examination under magnification showed no crush marks to the upper fitment.Functional and pressure testing was performed and a leak from the silicone tubing near the upper fitment was observed.The root cause of the leak was a tear in the silicone segment.The cause of the tear was not identified.
|
|
Event Description
|
The customer reported that after 1 min of infusing solumedrol, the user observed drops dripping from the pump module.The user opened the door and noted fluid squirting out of tubing.No patient harm was reported.
|
|
Search Alerts/Recalls
|