Model Number IBI-81202
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2017
Event Type  malfunction  
Manufacturer Narrative
The device was returned due to a signal issue.Electrodes 3-7 were distally displaced consistent with the electrode ring getting caught on the introducer sheath during the procedure and the reported signal issue.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the displaced electrode is consistent with damage during use.
 
Event Description
This report is to advise of an event of a dislodged electrode observed during analysis confirming the reported signal loss.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(AF-IRVINE)
2375 morse ave
irvine CA 92614
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7516743
MDR Text Key108502377
Report Number2030404-2018-00038
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K961924
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Model NumberIBI-81202
Device Catalogue NumberIBI-81202
Device Lot Number6084179
Other Device ID Number05414734303439
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial