Model Number ZA9003
Device Problems Device Operates Differently Than Expected (2913); Appropriate Term/Code Not Available (3191)
Patient Problem No Code Available (3191)
Event Date 04/16/2018
Event Type  Injury  
Manufacturer Narrative
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that the intraocular lens (iol) was implanted into patient¿s eye after a vitrectomy.Reportedly, the vitrectomy was done for other reasons.The account commented that after the iol was inserted into the patient's operative eye, the lens kept flipping over.The lens was removed and replaced with another lens of the same size with no problems.The incision enlargement was performed to remove the lens.No patient injury was reported.No further information provided.
 
Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no other complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Type of Device
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
Manufacturer (Section G)
JOHNSON & JOHNSON SURGICAL VISION, INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pam mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key7522081
MDR Text Key108515299
Report Number2648035-2018-00724
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474529083
UDI-Public(01)05050474529083(17)221018
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P990080
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date10/18/2022
Device Model NumberZA9003
Device Catalogue NumberZA90030235
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-