ETHICON INC. ETHILON* BLK 5-0 45CM (1)SC-20; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number 14501T
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2018 and suture was used.During the procedure, the needle pulled off.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
(b)(4).An overwrap, a paper folder, a thread and a loose needle were returned for analysis.During the visual inspection of the sample under magnification, the swage and attachment area was noted to be as expected and the thread still attached in the needle, however; the end was observed cut.The thread was inspected along of the strand and the end of the suture was noted cut (attachment region) that appear to be by use of a surgical instrument.According to the sample condition, no needle pull off was found, but it was noted breakage suture due to improper handling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
Manufacturer (Section G)
ETHICON INC.-BRAZIL
rodovia presidente dutra
km 154
sao paolo
BR  
Manufacturer Contact
darlene kyle
p.o. box 151, route 22 west
somerville, NJ 08876-0151
9082182792
MDR Report Key7522682
MDR Text Key108655404
Report Number2210968-2018-72877
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K946173
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Catalogue Number14501T
Device Lot NumberAJ6041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/24/2018
Initial Date FDA Received05/17/2018
Supplement Dates Manufacturer Received05/22/2018
Supplement Dates FDA Received06/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
-
-