A complaint history review and service history review for similar complaints was performed for hba1c calibrator lot zs7001 from the 20th of march 2017 through awareness date on (b)(6) 2018.There were no similar complaints identified during the searched period.A complaint history review and service history review for similar complaints was performed for serial (b)(4) from the 20th of march 2017 through awareness date on (b)(6) 2018.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 3, assay operations, under section 3.6 calibration states the following: we recommend calibrating the analyzer once a week.The laboratory may establish a longer calibration interval based on daily qc results.Be sure to calibrate in the following situations.If drift in qc is noted- the tosoh automated glycohemoglobin analyzer hlc-723g8 is a very precise instrument and you will note that the qc remains the same day after day.If there is a drift in qc values, it is necessary to recalibrate.When control values assayed are out of range calibrate when the control assay value falls outside the qc range.Measure the control sample again to confirm that it falls within the qc range before assaying a patient sample.After column replacement -calibrate after a new column has been installed.After analyzer maintenance - calibrate after periodic maintenance or repair.When an assay condition is being modified - calibrate when a parameter value (such as the flow factor) is being changed on the analyzer.The most probable cause of the reported issue was due to calibrators.(b)(4).
|