• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON THERMAGARD NEBULIZER HEATER; BREATHING SYSTEM HEATER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON THERMAGARD NEBULIZER HEATER; BREATHING SYSTEM HEATER Back to Search Results
Catalog Number 5705
Device Problems Overheating of Device (1437); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device involved in this complaint has not been received by the manufacturer at the time of this report.The investigation into this complaint is still in progress.
 
Event Description
Customer complaint alleges the unit over heated during a patient use.It was reported the patient is using a replacement device.There was no report of patient injury.
 
Manufacturer Narrative
(b)(4).The sample was returned and sent on the manufacturer (pegasus research corp).Pegasus reports that upon receipt of unit it was found that the heater was functional.The heater was tested for electrical safety and temperature output, and found to be within specifications.The unit was left on, with and without open case, for eight consecutive hours on each case, and the unit did not present variations in temperature out of the ones in specifications.The approximate age of the heater is 277 days.Based on the investigation performed, the complaint could not be confirmed.There were no issues found with the returned sample.
 
Event Description
Customer complaint alleges the unit over heated during a patient use.It was reported the patient is using a replacement device.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON THERMAGARD NEBULIZER HEATER
Type of Device
BREATHING SYSTEM HEATER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key7526916
MDR Text Key108692308
Report Number3011137372-2018-00136
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5705
Device Lot NumberSN:21910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/08/2018
Initial Date FDA Received05/18/2018
Supplement Dates Manufacturer Received06/25/2018
Supplement Dates FDA Received06/26/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-