ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problems Sticking (1597); Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/23/2018
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There is one other complaint in the lot.Additional information was requested.(b)(4).
 
Event Description
A nurse reported that during an intraocular lens (iol) implant procedure, one of the haptics stuck to the optic of the lens and the lens was torn.There was patient contact, but no patient harm.Additional information was requested.
 
Manufacturer Narrative
The device and the lens were returned separately.The plunger lock and the lens stop were removed.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger has been retracted to mid-nozzle.A broken haptic is observed.The haptic is on the left side of the plunger in the groove (correct).The distal portion is facing the end of the nozzle.The leading haptic and the optic were returned and were broken in a small bag.Viscoelastic was observed on the lens.A qualified viscoelastic was used.The root cause for the complaint could not be determined.The optic and lead haptic were returned separate and were broken.The broken trailing haptic is present in the device on the correct side of the plunger extended in the plunger groove.The distal portion is facing the end of the nozzle.The plunger was retracted.The plunger position in relation to the broken haptic during advancement cannot be determined.The haptic position may indicate the plunger was not fully advanced.If the plunger is not fully advanced the trailing haptic may not release properly from the device.Additional information was provided.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7527523
MDR Text Key108807038
Report Number1119421-2018-00628
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Model NumberAU00T0
Device Lot Number12556421
Other Device ID Number00380652394741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age43 YR
-
-