• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC LAP 5MM SHEAR 36CM; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC LAP 5MM SHEAR 36CM; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HAR36
Device Problem Failure to Power Up (1476)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/28/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # n92t72.Investigation summary: the analysis results found that the device was received with the black tissue pad detached and not returned but with evidence of body fluids and tissue pad material in the groove section of the clamp arm.The device was connected to a test handpiece and functionality tested on a gen11.The device was functional and worked as intended.The device was analyzed, and it was determined that it was connected in more than one generator.The device is intended and labeled for single patient use.If the device is connected to multiple generators an alert screen will be displayed indicating ¿adaptive tissue technology features are not available in this device¿.The device was disassembled to inspect the internal components and no anomalies were found.This device is packaged and sterilized for single use only.Multiple patient use may compromise the device integrity or create a risk of contamination that, in turn, may result in patient injury or illness.Based on the condition of the tissue pad, a probable cause for this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.No conclusion could be reached as to what caused this issue.A batch history record review was performed, and a nc was created during the manufacturing of this batch but was not related to the event description.Additionally, no defects or protocols related to the complaint were found during the manufacturing process.
 
Event Description
It was reported that during an unknown procedure, the device could not be activated during the procedure.Changed to another one to complete the procedure.There was no report on patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC LAP 5MM SHEAR 36CM
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7527774
MDR Text Key108867905
Report Number3005075853-2018-10009
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036014444
UDI-Public20705036014444
Combination Product (y/n)N
PMA/PMN Number
K120729
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Catalogue NumberHAR36
Device Lot NumberN92T72
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2018
Date Manufacturer Received04/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-