The engineering evaluation was completed and observations were: a viabahn device with a loaded endoprosthesis was returned.The distal tip was separate from the rest of the delivery catheter.The distal shaft, upon which the endoprosthesis is mounted, was exposed 0.2 cm on the transition end.The distal tip end of the endoprosthesis measured 1 cm.This portion of the distal shaft¿s pebax does not appear to be interrupted.The distal tip remained on a guidewire and a small white piece of material consistent with plastic was sticking out of the transition end of the distal tip.After the distal tip was taken off of the guidewire, the white plastic piece could be removed from the distal tip.The radiopaque marker band remained attached to the white piece when it was removed from the distal tip.The white plastic piece measured 1.3 cm in length and 1.1 cm from the end of the marker band to the end of the piece.The white plastic piece transitions to a black color about half way through the length.The plastic piece is curved, fitting a similar shape to the inside of the distal tip.The piece is about 150° of the circumference of the distal tip when looking at the cross sectional area.The distal tip appears to have been melted since there is a ¿bump¿ approximately 0.5 cm the tip end of the distal tip.The inner diameter dimensions of the tip appear to be within the acceptable range.The white plastic piece is not a component used by the distal tip supplier and they do not know its origin.There is nothing in the viabahn manufacturing process that resembles this material or color.Based on the device examination performed, no manufacturing anomalies were identified.
|