MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE TIBIAL TRIAL EXTRACTOR; ATTUNE INSTRUMENTS : EXTRACTION INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE TIBIAL TRIAL EXTRACTOR; ATTUNE INSTRUMENTS : EXTRACTION INSTRUMENTS Back to Search Results
Catalog Number 254500138
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); Not Applicable (3189)
Event Date 04/20/2018
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
During a tka the surgeon was sizing poly liners.As the surgeon was lifting the tibial poly extractor the right edge of the extractor broke off hitting the floor.The broken piece was not returned when it reached spd.Patient consequence? :no.
 
Manufacturer Narrative
Product complaint # pc-(b)(4).Investigation summary examination of the returned device confirms the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE TIBIAL TRIAL EXTRACTOR
Type of Device
ATTUNE INSTRUMENTS : EXTRACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork, munster
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
chad gibson
700 orthopaedic dr.
warsaw, IN 46581-0988
5743725905
MDR Report Key7529508
MDR Text Key108749072
Report Number1818910-2018-60082
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295131021
UDI-Public10603295131021
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500138
Device Lot NumberBFA0V4F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/20/2018
Initial Date FDA Received05/21/2018
Supplement Dates Manufacturer Received05/29/2018
05/29/2018
Supplement Dates FDA Received05/30/2018
05/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-