(b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the account to address the reported event.The fse inspected the instrument and found a leak at the sample loop and replaced it.Fse used the customer pipettes to reconstitute controls.Then, precision and controls were performed and passed within range.The control results were the following: level 1 5.9 (4.01-7.45) and level 2 9.3 (6.21-11.53).The life count was 49949.The g8 instrument was then operational in accordance with manufacturer's specifications.No further action was required by the fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: 6.3 error messages when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages with these errors, the assay stops and the analyzer immediately enters stand-by state.The 101 pressure low.The pressure will not rise because the pump is unable to run due to air bubbles in the pump check valve.If the elution buffer is empty, place a new elution buffer and execute reagent change.Next, execute drain flush.See "chapter 5 section 5.5: pump air removal".Execute manual pumping using the pump key in the main screen (second screen), and open and close the drain valve 2 or 3 times.If the pressure rises when the drain valve is closed, the operation is complete.If the pressure still does not rise or stabilize, execute drain flush again.In addition, confirm that the drain valve is securely closed.The most probable cause of the reported event was sample loop failure due to possible clot.
|
(b)(4), per exemption number e2017013.Submission of this report does not constitute an admission that the importer or manufacturer's product caused or contributed to the event.Device evaluation by manufacturer: the sample loop was received for evaluation on 23-may-2018.Visual inspection was performed which revealed damage to the part.The part was creased in the middle and one of the ferrules were intstalled incorrectly on the end of the loop.The reported product problem was confirmed.Evaluation codes: results: 3211 deformation problem, 3201 installation problem.Conclusion code: 14 device incorrectly prepared for use or modified.13 device difficult to operate.
|