• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHAMPION MANUFACTURING MECHANICAL CHAIR/TRANSPORT CHAIR; MECICAL RECLINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHAMPION MANUFACTURING MECHANICAL CHAIR/TRANSPORT CHAIR; MECICAL RECLINER Back to Search Results
Model Number 547FL06-TS
Device Problem Device Handling Problem (3265)
Patient Problem Laceration(s) (1946)
Event Date 02/18/2017
Event Type  Injury  
Manufacturer Narrative
Review of operator manual that accompanied the chair page 7 states proper in-service procedures.Pages 16 and 17 contain a caution that states: caution: this chair has moving parts that may create pinch points.Work carefully.Have your hands clear of the mechanism when you move the chair from one position to another.In addition to the initial shipment of chairs, there were four additional shipments of similar chairs received at this facility since this specific chair was purchased.All had operator manuals noting the proper operation of the recliners, including directions for staff to safely assist a patient in positioning the chair if they need assistance.The manual also contains cautions about keeping your hands clear of the mechanism.The in-service packet also comes with a video showing safe operation of the chair by a patient and by staff assisting the patient.Review of service parts and sales activity at this particular clinic shows no significant repair to the mechanical parts of their chairs.Routine follow up call noted chairs operating properly and there were no questions or concerns from staff.On 11 april 2018, champion verified with the facility that they no longer had serial (b)(4).The facility representative said he thought they discarded the chair when they purchased all new chairs from champion.
 
Event Description
A technician was assisting a patient onto a recliner.The actuator handle was pulled to release the footrest and the technician's hand was in the path of the opening footrest.The technician pinched his left ring finger in the mechanism of the chair, resulting in a skin trauma of laceration 2 cm square in size.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MECHANICAL CHAIR/TRANSPORT CHAIR
Type of Device
MECICAL RECLINER
Manufacturer (Section D)
CHAMPION MANUFACTURING
2601 industrial parkway
elkhart IN 46516
Manufacturer (Section G)
CHAMPION MANUFACTURING
2601 industrial parkway
elkhart IN 46516
Manufacturer Contact
christine miller
2601 industrial parkway
elkhart, IN 46516
MDR Report Key7533699
MDR Text Key108905360
Report Number1834066-2018-00001
Device Sequence Number1
Product Code KMN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Attorney
Type of Report Initial
Report Date 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Medical Technologist
Device Model Number547FL06-TS
Device Catalogue Number54 SERIES RECLINER
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/10/2018
Initial Date FDA Received05/22/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age67 YR
Patient Weight75
-
-