Catalog Number 1011493-18
Device Problems Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/15/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The instructions for use states: the rx herculink elite renal stent system is indicated for use in patients with atherosclerotic disease of the renal arteries following sub-optimal percutaneous transluminal renal angioplasty (ptra) of a de novo or restenotic atherosclerotic lesion (15 mm in length) located within 10 mm of the renal ostium and with a reference vessel diameter of 4.0 and 7.0 mm.The investigation determined that the reported difficulties were due to case circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the procedure was to treat a heavily calcified iliac artery.As a 5.0 x 18mm herculink elite stent delivery system was being advanced, the physician deemed it necessary to use a different size device.During removal, the delivery system got stuck on the guiding catheter (gc).The device was removed as a single unit including the gc.Another herculink was used to successfully complete the procedure.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Type of Device
RENAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7535748
MDR Text Key109005365
Report Number2024168-2018-03833
Device Sequence Number1
Product Code NIN
UDI-Device Identifier08717648078026
UDI-Public08717648078026
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number1011493-18
Device Lot Number7071961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/16/2018
Initial Date FDA Received05/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
Patient Weight112
-
-