DEPUY SYNTHES PRODUCTS LLC RADIOLUCENT-DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 511.300 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/07/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Reporter¿s phone number: (b)(6).Unique identifier (udi): (b)(4).Device manufacture date was unknown.This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the radiolucent drive device coupling on the tool side was worn and the drill receptive came out.It was further determined that the device failed pretest for check the untrue running, check the machine coupling and check the tool coupling.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|