Model Number MMT-754WWS
Device Problems Erratic or Intermittent Display (1182); Moisture Damage (1405)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/07/2018
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported that the insulin pump was exposed to moisture and the screen was difficult to read.Customer's blood glucose value was unknown.Insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device received with blank display due to moisture damaged lcd board during visual inspection.Unable to perform operating current test, unexpected restart error test, self test and displacement test due to blank display.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWS
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7543637
MDR Text Key109413134
Report Number3004209178-2018-73887
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169507685
UDI-Public(01)00643169507685
Combination Product (y/n)N
Reporter Country CodeNZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-754WWS
Device Catalogue NumberMMT-754WWS
Device Lot NumberB5754WWSJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/07/2018
Initial Date FDA Received05/25/2018
Supplement Dates Manufacturer Received08/14/2018
Supplement Dates FDA Received09/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-