The complaint was dispatched to the field service engineer (fse) department for further investigation.Fse conducted follow-up with the customer over-the-phone to address the reported event.The fse advised the customer to clean the shaft and screw drive for the z1 axis with alchohol.Then the fse instructed the customer on removing the needle for cleaning.After cleaning the shaft and screw drive the fse instructed the customer on how to lube the shaft and screw drive with triflow.The customer ran more than twenty samples after the re-assembly of the needle.The run went to completion without any errors.Instrument is functioning as intended and was released.No further action is required by the fse.A complaint history review and service history review for similar complaints was performed for serial number (b)(4) from the (b)(6) 2017 through awareness date on the (b)(6) 2018.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: 710 z1-axis error explanation: operation error in z1-axis.Countermeasure: inspect z1-axis.Execute smp.Reset.The most probable cause of the reported issue was due to the sample z axis requiring lubrication.(b)(4).
|