MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number MCP00703323
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/14/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).Field service technician has been sent for investigation.The issue was already solved by the hospital after spinning pump head manually and resetting it on the pump.No parts have been replaced.Failure could be confirmed.Most probable root cause could be determined as incomplete/ faulty setting of the pump (service issue).Thus no product related malfunction could be confirmed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Correction:according to service report, the issue was solved after spinning pump head manually and resetting of the pump.Functional tests have been performed acc.To service manual.Device passed all tests.Corrective action: since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.
 
Event Description
According to complaint report, received on 2018-05-15: returned a tpm from repair of "rev.Relay" error, when put on head it showed error safety-s.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HL-20 INTEGRATED PERFUSION SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key7549107
MDR Text Key109626859
Report Number8010762-2018-00186
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCP00703323
Device Catalogue Number701035075
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/15/2018
Initial Date FDA Received05/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-