(b)(4).Device evaluation by manufacturer: a 13-month complaint history review and service history review for similar complaints was performed for serial (b)(4) from 14-apr-2017 through aware date (b)(6) 2018.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.The 100 pressure high: the pump pressure exceeded the upper limit (15 mpa) set in the pres high parameters.When the filter or column replacement period has been exceeded, first replace the filter or column.If the pressure is still high, remove the inlet and outlet flow line around the column and filter, and determine which part is the cause of the high pressure.Then, contact a technical support representative.If the pressure displayed on the screen is: (a) greater than the pressure on the column inspection report + 4 mpa, then replace the filter.(b) less than the pressure on the column inspection report, then proceed with priming the column.The most probable cause of the reported event was due to a failure or fault of column.
|
Customer reported 100 pressure high error on start up with the g8 instrument.The customer stated that they changed the filter and the pressure remained high.The column count was 1900 injections and as a result, the customer flipped the column but the pressure was still high.All elution buffer bags were full and the waste had been emptied and was flowing without issue.Technical support (ts) had the customer remove the tubing from the right side of the column and check the pressure, which dropped to 0.5 mpa.Tss sent the customer a replacement column.When the customer received the column and placed it, the high pressure error resolved.Customer reported that after the column was installed, the pressure was 8.5 mpa.Customer then calibrated and ran controls with acceptable results.The instrument was operational.The customer did not have a replacement column and therefore was sent one by technical support, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|