DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 05.001.201 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/11/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.(b)(4).The manufacturing location was unknown.Device manufacture date: the device manufacture date is unavailable.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
This is report 1 of 2 for the same event.It was reported from (b)(6) that the handpiece device and the chuck with key device were overheating when used together.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The manufacturer location was documented as unknown in the initial medwatch report.The location has been updated to (b)(4).Contact office name/address have been updated accordingly to reflect the corrected manufacturing facility.It was documented in the initial medwatch report that the date of manufacture (dom) was unknown.The dom (mar 19, 2009) has been updated to reflect the date the device was manufactured.The unique identifier (udi) has been updated accordingly.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The handpiece device was evaluated and it was determined that the coupling tool side was not functioning and was defective.It was further determined that the device failed pretest for check the machine coupling and check for leakage.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.It was documented in the initial report that the serial number was unknown.It has been updated as (b)(4).Udi has been updated accordingly.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|