Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems No Code Available (3191); Pericardial Effusion (3271)
Event Date 05/08/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure, a pericardial effusion was observed.A pericardiocentesis was performed and the patients vitals signs remained stable.The case was aborted and the patient was under general anesthesia.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Event summary: the patient data files did show that during multiple injections there was oscillation within the flow and temperature performance with catheter 2af234 / 50663 which was used for nine injections.A clinical issue was encountered during cryo ablation procedure, and a pericardial effusion was noted.The sheath was not returned for investigation.In conclusion, a clinical issue (pericardial effusion) was encountered post procedure.The case was aborted under while the patient was under general anesthesia.There is no indication of product malfunction related to the adverse event.There was no product returned for investigation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7554299
MDR Text Key109579089
Report Number3002648230-2018-00345
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2AF234 BALLOON CATHETER
Patient Outcome(s) Required Intervention;
Patient Age74 YR
-
-