• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/30/2018
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by manufacturer: a field service engineer (fse) followed-up with the customer over-the-phone and confirmed that the issue was resolved by replacing a bent sampling needle assembly.The customer ran quality controls and patient samples without any issues.The g8 instrument was performing according to manufacturer's specifications.No further action was required by the fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from (b)(4) 2017 through aware date (b)(4) 2018.There were no other similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: 6.3 error messages when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.The 706 syringe-l error: explanation: operation error in syringe-l.Countermeasure: inspect x1-axis.Inspect syringe-l.Execute smp.Reset.Chapter 5, maintenance procedures, under section 5.10 provides step-by-step instructions on the sampling needle assembly replacement.The most probable cause of the reported 706 syringe-l error message was due to a bent sampling needle assembly.Additional manufacturer narrative: (b)(4).
 
Event Description
A customer reported getting 706 syringe-l error message with the g8 instrument.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 105-8 623
JA  105-8623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7555565
MDR Text Key109767163
Report Number8031673-2018-00447
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA05/30/2018
Distributor Facility Aware Date04/30/2018
Device Age6 YR
Event Location Hospital
Date Report to Manufacturer05/30/2018
Date Manufacturer Received04/30/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-