• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING LIBERTY SELECT CYCLER ASSY(NON-VALUATED); SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Model Number LIBERTY SELECT CYCLER
Device Problems Overfill (2404); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/11/2018
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the plant¿s investigation.
 
Event Description
A peritoneal dialysis patient contacted technical services with ultra-filtration questions.Treatment data was provided.Treatment data was assessed for 150% and 180% iipv and did reveal an episode of increased intraperitoneal volume (iipv) on (b)(6) 2018.The largest drain volume from this treatment occurred during drain 4 where 3823ml drained.This drain is 191% of the prescribed fill volume.In order to be a reportable malfunction, at least 180% of the prescribed fill volume would have to have been drained.A 180% of fill volume 2002ml is 3603.Therefore, a reportable malfunction has occurred.The patient was advised to discontinue use of the cycler.The cycler was replaced.Additional information was solicited but unavailable.
 
Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An external visual inspection was performed with no physical damage noted.A simulated therapy was initiated and completed on the cycler without complication.The weighed fill volumes were found to be within tolerance and fill times were within specification.The cycler underwent system air leak and valve actuation testing and was found to meet product specifications.A patient sensor calibration test was also performed and the cycler was found to be within tolerance.Load cell verification testing was performed with no issues noted.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, there were no malfunctions that could have caused or contributed to the reported event.The cycler performed as designed and an associated cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key7556143
MDR Text Key109621979
Report Number2937457-2018-01505
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
PMA/PMN Number
K171652
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLIBERTY SELECT CYCLER
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2018
Device AgeMO
Date Manufacturer Received06/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DELFLEX PD SOLUTION; LIBERTY CYCLER SET; DELFLEX PD SOLUTION; LIBERTY CYCLER SET
-
-